12" Patient Monitor ICU CCU Vital Sign For Hospital ECG NIBP RESP TEMP SPO2 PR (PM 12F )Features:
1. Unique accessory management cabinet
2. 12.1-inch high resolution color TFT display
3. USB interfaces support easy software upgrade and data transfer
4. Three Working Modes: Monitoring, Surgery and Diagnosis
5. Simple and friendly operating display interface.
6. Multi-display modes suitable for different applications:
Standard Interface, Large font, ECG Standard Full Display, OXY, Trend Table, BP Trend.
7. Built-in high-capacity battery for up to 3 hours of continuous work
Standard Parameters:
ECG, RESP, TEMP, NIBP, SPO2, PR
Technique Specifications:
ECG
1. Lead Mode: 5 Leads (I, II, III, AVR, AVL, AVF, V)
2. Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
3. Heart Rate: 15-300 BPM (Adult); 15-350 BPM (Neonatal)
① Resolution: 1 BPM
② Accuracy: ±1%
③ Sensitivity >200 uV(Peak to peak)
4. ST measurement range: -2.0 ~ +2.0 mV
① Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, which is greater
Other Range: unspecified
5. Sweep speed: 12.5 mm/s, 25mm/s, 50mm/s
6. Bandwidth:
7. Diagnostic: 0.05~130 Hz
8. Monitor: 0.5~40 Hz
9. Surgery: 1~20 Hz
SPO2
1. Measuring Range: 0 ~ 100 %
2. Resolution: 1%
3. Accuracy: 70% ~ 100% (±2 %)
4. Pulse Rate: 20-250 BPM
5. Resolution: 1 BPM
6. Accuracy: 3 BPM
NIBP
1. Method: Oscillometry
2. Measure mode: Manual, Auto, STAT
3. Unit: mmHg, kPa
4. Measure and alarm range:
5.1 Adult Mode
① SYS 40 ~ 280 mmHg
② DIA 10 ~ 225 mmHg
③ MEAN 20 ~ 240 mmHg
5.2 Pediatric Mode
① SYS 40 ~ 220 mmHg
② DIA 10 ~ 160 mmHg
③ MEAN 20 ~ 170 mmHg
5.3 Neonatal Mode
① SYS 40 ~ 135 mmHg
② DIA 10 ~ 100 mmHg
③ MEAN 20 ~ 110 mmHg
6. Resolution: 1mmHg
7. Accuracy: 5mmHg
RESP
1. Method: Impedance between RA-LL or RALA
2. Measurement Range:
① Adult: 10 ~ 120 BrPM
② Neonatal / Pediatric: 7 ~ 150 BrPM
3. Resolution: 1 BrPM
4. Accuracy: 2 BrPM
TEMP
1. Measure and Alarm Range: 0 ~ 50℃
2. Resolution: 0.1℃
3. Accuracy: 0.1℃
Standard Configurations:
Main Unit *1
5-lead ECG Cable *1
Disposable ECG Electrode *5
Adult Spo2 Probe *1
Adult NIBP Cuff *1
NIBP Extension Tube *1
Temperature Probe *1
Power Cable *1
User Manual *1
Fuse Patient Monitor ICU CCU Vital Sign For Hospital ECG NIBP RESP TEMP SPO2 PR (SNP900N )Features:
1. 12.1" TFT color LCD,800*600dpi
2. Standard 6 para: ECG, SpO2, NIBP, TEMP, RESP, PR
3. Fanless design, quiet, energy-saving and clean
4. With the standard interface, OxyCRG, trend chard, big character interface and view bed,
5. Operation mode: keys and buttons.
6. Display up to 8-channel waveform on one screen.
7. NIBP measurement mode: Manual/AUTO/Continuous, storage for 1000-group NlBP data
8. Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
9. AC/DC, built-in rechargeable lithium battery achieves uninterrupted monitoring.
Standard Parameters:
ECG, RESP, TEMP, NIBP, SPO2, PR
Technique Specifications:
ECG
1. Lead Mode: 5 Leads (I, II, III, AVR, AVL, AVF, V)
2. Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
3. Heart Rate: 15-300 BPM (Adult); 15-350 BPM (Neonatal)
① Resolution: 1 BPM
② Accuracy: ±1%
③ Sensitivity >200 uV(Peak to peak)
4. ST measurement range: -2.0 ~ +2.0 mV
① Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, which is greater
Other Range: unspecified
5. Sweep speed: 12.5 mm/s, 25mm/s, 50mm/s
6. Diagnostic: 0.05~130 Hz
7. Monitor: 0.5~40 Hz
8. Surgery: 1~20 Hz
SPO2
1. Measuring Range:35%~100%
2. Resolution: 1%
3. Certainty of measurement: In the range of 70% to 100%, the precision should be ± 3%
4. Pulse Rate: 20-250 BPM
5. Resolution ratio: 1 BPM
6. Accuracy: 2 BPM
NIBP
1. Method: Pulse wave oscillation method
2. Measure mode: Manual / Automatic / Continuous
3. Unit: mmHg, kPa
4. Measure and alarm range:
5.1 Adult Mode
① SYS 30 ~ 255 mmHg
② DIA 10 ~ 220 mmHg
③ MEAN 20 ~ 235 mmHg
5.2 Pediatric Mode
① SYS 30 ~ 255 mmHg
② DIA 10 ~ 220 mmHg
③ MEAN 20 ~ 235 mmHg
5.3 Neonatal Mode
① SYS 30 ~ 135 mmHg
② DIA 10 ~ 110 mmHg
③ MEAN 20 ~ 125 mmHg
6. Resolution ratio: 1mmHg
7. Accuracy: 5mmHg
RESP
1. Method: Impedance between RA-LL or RALA
2. Measurement Range:
① Adult: 0 ~ 120 BrPM
② Neonatal / Pediatric: 0 ~ 150 BrPM
3. Resolution ratio: 1 BrPM
4. Accuracy: 2 BrPM
TEMP
1. Measure and Alarm Range: 0 ~ 50℃
2. Resolution ratio: 0.1℃
3. Certainty of measurement: ± 0.1℃ (without probe error)
Standard Configurations:
Patient Monitor *1
Power Cord *1
Spo2 Sensor *1
Temp Sensor *1
5-lead ECG Cable *1
Electrodes *10
NIBP Cuff *1
NIBP Extension Hose *1
User Manual *1
Ground Line />12 inch patient monitor with ETCO2 (No additional purchase required) ECG/NIBP/RESP/TEMP/SPO2/PR/ETCO2
PM12F
Features:
1.Unique accessory management cabinet.
2.12.1 inch high resolution color TFT display.
3.Includes a ETCO2 sensor module; no additional purchase required.
4.Three Working Modes: Monitoring, Surgery and Diagnosis.
5.Simple and friendly operating display interface.
6.Multi-display modes suitable for different applications:
Standard Interface, Large font, ECG Standard Full Display, OXY , Trend Table, BP Trend.
7.Built-in high capacity battery up to 3 hours of continuous work
Standard Parameters:
ECG, RESP, TEMP, NIBP, SPO2, PR, ETCO2
Technique Specification:
ETCO2
1.Monitoring range: 0% ~ 13%
2.Resolution: 1 mmHg
3.Accuracy: 2 mmHg @ < 5.0% CO₂ (at ATPS)
4.Respiration Rate: 3 - 150 bpm
ECG
1.Lead Mode: 5 Leads (I, II, III, AVR, AVL,AVF, V)
2.Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
3.Heart Rate: 15-300 BPM (Adult); 15-350 BPM (Neonatal)
① Resolution: 1 BPM
② Accuracy: ±1%
③ Sensitivity >200 uV(Peak to peak)
4.ST measurement range: -2.0 ~ +2.0 mV
① Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, which is greater
Other Range: unspecified
5.Sweep speed: 12.5 mm/s, 25mm/s, 50mm/s
6.Bandwidth:
7.Diagnostic: 0.05~130 Hz
8.Monitor: 0.5~40 Hz
9.Surgery: 1~20 Hz
SPO2
1.Measuring Range: 0 ~ 100 %
2.Resolution: 1%
3.Accuracy: 70% ~ 100% (±2 %)
4.Pulse Rate: 20-250 BPM
5.Resolution: 1 BPM
6.Accuracy: 3 BPM
NIBP
1.Method: Oscillometry
2.Measure mode: Manual, Auto, STAT
3.Unit: mmHg, kPa
4.Measure and alarm range:
5.1 Adult Mode
① SYS: 40 ~ 280 mmHg
② DIA: 10 ~ 225 mmHg
③ MEAN: 20 ~ 240 mmHg
5.2 Pediatric Mode
① SYS: 40 ~ 220 mmHg
② DIA: 10 ~ 160 mmHg
③ MEAN: 20 ~ 170 mmHg
5.3 Neonatal Mode
① SYS: 40 ~ 135 mmHg
② DIA: 10 ~ 100 mmHg
③ MEAN: 20 ~ 110 mmHg
6.Resolution: 1mmHg
7.Accuracy: 5mmHg
RESP
1.Method: Impedance between RA-LL or RALA
2.Measurement Range:
① Adult: 10 ~ 120 BrPM
② Neonatal / Pediatric: 7 ~ 150 BrPM
3.Resolution: 1 BrPM
4.Accuracy: 2 BrPM
TEMP
1.Measure and Alarm Range: 0 ~ 50℃
2.Resolution: 0.1℃
3.Accuracy: 0.1℃
Standard Configuration:
Main Unit *1
ETCO2 Sensor *1
5-lead ECG cable *1
Disposable ECG Electrode *5
Adult Spo2 probe *1
Adult NIBP cuff *1
NIBP extension tube *1
Temperature probe *1
Power Cable *1
User Manual *1
Fuse *3FDAStatement: